INTERFERÊNCIA ELETROMAGNÉTICA CAUSADA POR EMISSÕES CONDUZIDAS EM ELETROMÉDICO

  • Danillo Fernandes dos Santos Padin Centro Universitário Brazcubas
  • André Amancio Almeida Centro Universitário Brazcubas
  • Antônio Carlos da Cunha Migliano Centro Universitário Brazcubas
  • Fernanda Talhati Centro Universitário Brazcubas
Palavras-chave: Cardiotocógrafo, Conformidade, Eletromédico, Emissões conduzidas

Resumo

RESUMO

A eficiência e efetividade do atendimento em clínicas e hospitais dependem dos equipamentos disponíveis para colaborar no diagnóstico, tratamento ou monitoramento de seus pacientes. Assim, espera-se que esses estabelecimentos de saúde tenham todos seus equipamentos em conformidade com as normas de segurança e registrados junto a Agência Nacional de Vigilância Sanitária (ANVISA). Neste sentido, tais exigências são previstas na NBR IEC 60601-2, que estabelece os níveis máximos de tensão para as emissões conduzidas, geradas pelos equipamentos eletromédicos. A avaliação de conformidade desses equipamentos torna-se necessária para garantir sua comercialização e, neste sentido, este trabalho tem por objetivo analisar um Cardiotocógrafo a partir de ensaios de emissões conduzidas, seguindo as normas FCC part 15(Federal Communications Commission), considerando os níveis máximos exigidos para equipamentos de classe B, trabalhando em faixas de frequências com níveis de tensões reduzidas, comparadas aos eletromédicos classificados na classe A. Desta forma, este estudo se restringiu a faixa de frequência de 0,15 a 30 MHz, com limite máximo de tensão de 46 dBuV. O ensaio foi realizado com um arranjo experimental constituído de um analisador de espectro, redes de impedâncias (LISN) e um microcomputador para o controle e automação do ensaio via interface IEE488, com o software de EMC/EMI E7415A. Foram obtidos valores de picos de tensão entre 46 e 56 dBuV, o que atestou a não conformidade do equipamento, principalmente nas regiões limítrofes de frequência. A proposta de saneamento dessas não conformidades envolveu a inserção de filtros de linha no cabo de alimentação do eletromédico. No entanto, essa solução não se tornou satisfatória para frequências menores que 1 MHz, por utilizar transdutores dos filtros com absorção em mais altas frequência.

Palavras chave: Cardiotocógrafo, Conformidade, Eletromédico, Emissões conduzidas.

ABSTRACT

The efficiency and effectiveness of care in clinics and hospitals depend on the equipment available to collaborate in the diagnosis, treatment or monitoring of their patients. Thus, it is expected that these health facilities will have all their equipment in compliance with safety standards and registered with the National Agency of Sanitary Surveillance (ANVISA). In this sense, such requirements are foreseen in NBR IEC 60601-2, which establishes the maximum voltage levels for the conducted emissions, generated by electromedical equipment. The conformity assessment of these equipments becomes necessary to guarantee their commercialization and, in this sense, this work has the objective to analyze a Actocardiotocograph from the tests of conducted emissions, following the rules FCC part 15 (Federal Communications Commission), considering the levels class B equipment, working in frequency bands with reduced voltage levels, compared to electromedicals classified in Class A. In this way, this study restricted the frequency range of 0.15 to 30 MHz, with a maximum limit of voltage of 46 dBuV. The assay was performed with an experimental arrangement consisting of a spectrum analyzer, impedance networks (LISN) and a microcomputer for the control and automation of the IEE488 interface test with the EMC / EMI software E7415A. Voltage peak values ​​were obtained between 46 and 56 dBuV, which demonstrated the nonconformity of the equipment, especially in the frequency boundary regions. The proposal to sanitize these nonconformities involved the insertion of line filters in the electromedical power cable. However, this solution did not become satisfactory for frequencies lower than 1 MHz, because it uses transducers of the filters with absorption at higher frequency.

Keywords: Actocardiotocograph, conducted emissions, electromedical equipment, frequency.

Biografia do Autor

{$author}, Centro Universitário Brazcubas

Graduando em Engenharia Elétrica da Braz Cubas Educação.

{$author}, Centro Universitário Brazcubas

Graduando em Engenharia Elétrica da Braz Cubas Educação.

{$author}, Centro Universitário Brazcubas

Professor Titular Doutor da Braz Cubas Educação e Pesquisador Titular Doutor do Instituto de Estudo Avançados, São José dos Campos, SP.

{$author}, Centro Universitário Brazcubas

Professora Assistente Titular, Braz Cubas Educação.

Publicado
2019-06-03

##plugins.generic.recommendByAuthor.heading##